To ensure the optimal use of your Sanibel disposable tab or snap electrodes, please observe the following:

  • Ensure that the skin is clean and dry before placing the electrode
    o It is recommended to use an alcohol wipe or similar to clean the newborn skin,

    especially if the skin is greasy or oily, prior to placing the electrodes

  • When applying electrodes to the skin, press down firmly on the entire electrode surface to improve adhesion

  • Use a small amount of conductive electrode gel to improve impedances, if required

  • Use pinch or button cables for snap electrodes and alligator cables for tab electrodes

  • Electrode adhesion time is limited to a maximum of 1 hour

  • Store unused electrodes in unopened pouches at temperatures 5°-27° C / 40°-80° F

  • Use before the expiry date printed on the pouch