Description: The following document outlines how to replace the attenuator covers on the AC40 and AD629. STEP 1: Remove both covers from the attenuators – do not use tools. STEP 2: Check cover revisi…
Description: The following document outlines how to replace the attenuator covers on the AC40 and AD629. STEP 1 Remove both the covers for the attenuators – do not use tools STEP 2 Check cover revisi…
Description: AC40, AD226, AD629 & NEW AT235 supports direct print-out to a range of standard PC USB printers with standard printer language HP PCL 5/HP PCL 3, e.g. HP and Canon printers. The instrume…
Description: When calibrating free field in AD528 in a 1 speaker setup, the calibration changes made in the software does not pull through and the calibration will be invalid. This is a fault in the…
Description: This document describes the components needed to remote deploy the Affinity/Equinox suite. This document is intended for IT persons specialized in remote deployment (i.e. SCCM) of applic…
Description: In Diagnostic Suite 2.9 it has been discovered that there can be some issues regarding transfer of data from the MT10 with the IR dongle. When trying to transfer data, the suite can cras…
Description: Because of changes in the compatibility between the Titan Display and the Keyboard PCA, both parts must be changed at the same time in Titans with serial number before SN0965669. This me…
Description: We have become aware of a problem with OAE data transfer from Titan handheld unit via Bluetooth, and we are presently working on a solution. Workaround: Meanwhile, the workaround to tran…
Purpose: To maintain a stable connection between the Titan and the clinical extension cable. Description: Keep the connectors on the Titan and the clinical extension cable/probe clean with a cotton b…
Description: Affinity cabinet has 8 screws that can become loose and thereby cause the lid to not operate properly. Purpose: Secure that all 8 screws are fixed. Solutions: Unscrew the screw; apply Lo…
Description: We have in some cases seen that ABRIS is not able to open in neither standalone mode nor trough OtoAccess®. The problem is related to backup files that are not deleted automatically. The…
Purpose: This technical note explains how to fix a Callisto™ when it does not emit any sound. Before following the instructions in this technical note, it should be verified that the firmware and the…