Tone Decay

Description
This is a test to help identify the adaptation of the auditory system (Carhart, 1957). It involves measuring the perceptual reduction in a continuous tone over time. This can indicate towards a cochlear or neural cause ofdeafness.
The test involves looking at the patient’s response to the onset of a supra-threshold sound and then their continuous response to this as it continues over time. For example, in Meniere’s disease this is detected correctly on onset but rapidly deteriorates due to dysfunctional hair cells (Carhart, 1957). A normal response should be maintained for a minute of stimulation. Should a patient not be able to maintain this, the stimulus intensity is increased until a minute is achieved. This is only increased up to a maximum 30dB suprathreshold.

Required Items

  • The AC40 hardware
  • Calibrated headphones or insert phones
  • Patient response button

Tone Procedure
The tone decay can be run on the AC40 as standalone or by using the AC40 with Diagnostic Suite.

  1. The patient’s audiometry is obtained.
  2. Press the Tests  button and select Tone decay  from the list of test by rotating the wheel.
  3. The patient is then instructed to continuously respond to the tone by pressing the response button if they hear the tone and not respond as the signal fades/is absent by letting go of the response button.
  4. Press Start.
    The test is administered with a pure tone presented at 5dB SL and then ascended in 5dB steps without interruption until the subject responds. As soon as the subject responds, the system will begin timing.
    If the tone is heard for a full one minute, the test is stopped.
    If the subject indicates that the patient no longer hears the tone before the minute criterion is reached, the intensity of tone is increased by 5dB without interrupting the tone, but the timing at the top of the screen is reset. The tone is continued to be raised in 5dB steps until an intensity is reached that allows the subject to perceive the tone for a full minute.
    The results are stored in the bottom of the test window for the given ear and frequency, showing the amount of decay occurring.
  5. As a time saving measure, Carhart (1957) suggested that the test should be terminated when the subject fails to respond 30dB above threshold.