Prepare the equipment
Test environment
The ideal test environment is a quiet room where lights and other electronic equipment are turned off.
Prepare the Infant
Patient state |
The infant should be sleeping or in quiet relaxed state. Sucking, blinking, crying or movement may affect testing. |
Skin preparation |
Typically, no skin preparation is required. |
Place electrodes |
Place surface electrodes using the desired montage (mastoid or nape). |
Connect cables |
Connect the cables from the PreAmplifier to the respective surface electrodes. |
Check impedance
Impedances are indicated by the green/amber dots on the infant as well as in numerical format on screen.
When the dot is amber, this means the impedance is poor (> 40kΩ). In this instance, it may be necessary to clean the skin and replace the surface electrode.
NB. Testing is possible when impedances are poor. This may, however, affect test time and measurements.
Place transducer/s
Place the probe or insert earphones in the infant’s ear/s or place the headset over the infant’s ears.
The probe light will turn green when a good seal is obtained.
Run test
Press the START button on the Titan or press the preamplifier button.
Patient noise (EEG)
After starting the test, the patient noise/EEG bar is displayed at the top of the screen.
If the patient noise goes above the black line, try to calm the infant to reduce movement, crying, sucking etc.
If the infant appears calm and the patient noise is not ideal, stop the test and select a protocol with a higher rejection level (if available).
If the patient noise/EEG signal is still not ideal, refer to the EEG troubleshooting section below.
Results
PASS result![]() |
REFER result![]() |
PASS – when the result reaches 100% for the pass criteria, PASS is displayed in green above the measurement.
REFER – when the result does not reach 100% for the pass criteria within the measurement time, REFER is displayed in amber above the measurement.
INCOMPLETE – if the test is stopped before a PASS or REFER is generated by the system, INCOMPLETE is displayed above the measurement indicating that the full test was not completed.
EEG troubleshooting
When the patient noise/EEG signal is poor and the patient is calm, try the following: