Cleaning the Titan Probe Tip

Probe, Short Extension Cable, Preamplifier

 

Clinical Extension Cable

1. Unscrew the probe cap.

 

1. Unscrew the probe cap. 

2. Remove the probe tip.

 

2. Remove the probe tip. 

3. Thread the stiff end of the cleaning floss into one of the tubes.

 

3. Thread the stiff end of the cleaning floss into one of the tubes. 

4. Pull the cleaning floss completely through the probe tip tube. Clean each of the tubes as required. Discard floss after use.

 

4. Pull the cleaning floss completely through the probe tip tube. Clean each of the tubes as required. Discard floss after use. 

5. Assemble the probe again.

 

5. Assemble the probe again. 


Notice
:
 
Do not use the cleaning tool to clean the probe base. This will destroy the filters.