Monothermal Warm Caloric Screening Test

Thursday, 19 November 2009 06:50
Published in Balance Assessment
Read 372 times

In some test situations, caloric responses may be available only for one temperature; either by design or by necessity:

  • Design: Monothermal screening test
  • Necessity: Patient may not be able to tolerate complete bithermal caloric procedure

The Interacoustics FireWire VNG systems VN415 and VO425 now offer the Monothermal Warm Caloric Screening Test with a modified equation for Unilateral Weakness, UW%

The Monothermal Warm Caloric Screening Test uses the first two warm caloric irrigations to calculate a screening value for caloric weakness called MWST.

If the MWST value is below 15% you can omit the cold irrigations, thereby shortening the test time by 15 minutes and exposing the patient to less testing.

MWST can be calculated with compensation for spontaneous nystagmus.