Interacoustics

AT235 Hardware Specifications

Basic facts

Size Length: 16.5 inches (42 cm)
Width: 12.5 inches (32 cm)
Height: 4 inches (10.2 cm)
Weight Weight: 9 pounds (4.1 kg)
Compatible Windows software Interacoustics OtoAccess™ Database.
Diagnostic Suite for PC-integration
(OtoAccess™ and NOAH compatible)
 

Probe tone

  • Frequency: 226Hz.
  • AT235h also: 678Hz, 800Hz, 1000Hz for traditional IYI-curve tympanometry. Level: 85 db SPL. Gain Control: AGC.

Air pressure

  • Control: Automatic.
  • Range: Default +200 to –400daPa (max. +300 to -600daPa).
  • Safety limitation: -800 daPa and +600 daPa.
  • Pressure change rate: Selectable in the set-up between 50, 150 and >250 daPa/s or automatic.
  • Compliance: Range: 0.1 to 6.0 ml (numerical: 0.1 to 8.0 ml).

Function

  • Automatic, where pump speed, start and stop pressure can be userprogrammed in the set-up.
  • AT235h: Auto and manual pump functions.5 dB

Medical CE-mark

  • Yes

Eustachian Tube Function

  • AT235: Function test for use when the eardrum is intact.
  • AT235h: Function tests for use with both intact and perforated eardrums.

Memory

Internal memory for two ears. Each ear: 6 Ipsi and 6 Contra recordings. Each may have up to 6 stimuli. Also, there is memory for additional manual reflex recordings. (Total max. 78 reflexes per ear).

Acoustic Reflex tests

  • Automatic reflex: Two independent user selectable protocols. Series of fixed intensities available.
  • Automated intensity search functions available for threshold search and reflex growth indication.
  • Free mixing of Ipsi and Contra.
  • Manual reflex: Manual control of all stimuli. May also be used to redo part of automated test results.
  • Reflex decay: Manual control, with stimulus duration of 10 sec. Ipsi or contralateral stimulation.

Frequencies and intensity ranges

  • Ipsilateral: Intensity up to 110dBHL.
  • Frequency: 250, 500, 1000, 2000, 3000, 4000, WB, HP, LP noise Contralateral: Intensity up to 120dBHL.
  • Frequency: 125 to 8000Hz, WB, LP, HP noise
  • Audiometry: Intensity from -10 to 120dBHL. Frequency from 125Hz to 8000Hz.

Attenuator

  • 1dB or 5dB steps.

Audiometer Functions

  • Manual Audiometry.
  • Automatic Audiometry according to ISO 8253-1 (Patient controlled Hughson-Westlake).

Printer

  • Built-in fast thermal printer with paper width: 112 mm.

Included in the box

  • ATP-AT235u Universal probe system with shoulder strap and wrist strap
  • HBC39 Single contralateral headset
  • Power cable (110/220V, please specify)
  • BET55 Ear tips in box
  • TPR26 3 Rolls of recording paper
  • Operation manual CD
  • Multilingual CE - instructions for use

Optional Parts

  • TDH39 Audiometric headset
  • EAR-Tone5A Insert phones for audiometry
  • Soft carrying bag
  • CIR33 contralateral insert phone
  • APS3 Response switch
  • CAT50 Calibration unit 0.2-0.5-2.0-5.0 ml
  • IES-2 Impedance ear simulator
  • OtoAccess™ Database
  • Diagnostic Suite for PC-integration (OtoAccess™ and Noah compatible)

Printer Option

  • AT235-xp is identical to the AT235 but has no built-in printer.
  • Suitable for installations where computer connection takes care of printing.

Impedance standard

  • IEC 60645-5/ANSI S3.39, Type 2.

Audiometer standard

  • IEC 60645-1/ANSI S.3.6, Type 4 Tone.

Safety standard

  • IEC 60601-1, Class I, Type B.

Consumption

  • 15VA, max. 45VA.
 

Computer Integration

Genral computer requirements
  • 1.2 GHz Pentium 4 class CPU
  • 512 MB ram
  • 1 GB available disk space
  • CD-Rom drive
  • USB slot
Display requirements
  • 1024x768 resolution
  • Hardware accelerated DirectX/Direct3D graphics card.
Software requirements
  • Windows XP (SP3), Windows 7 (SP1)

Additional service and support

To find additional support please call us or go to our support center

You are here: Home AT235 Specifications
24 May 2013

Tympanometers

Titan software modules

Wideband Tympanometry (WBT) from Interacoustics explores middle ear diagnostics from a whole new perspective. Explore WBT

Do you want to know more about the Titan hardware? Explore Titan

Support

Go to the main e-iconSupport section

Events