Interacoustics

Technical Specifications

Basic facts

Size Lenght: 20.6 inches (52.2 cm)
Width: 14.4 inches (36.6 cm)
Height: 3.9 inches (9.8 cm)
Weight 17.4lb (7.9 kg)
Power supply Built-in power supply,
100-240 V, 50-60Hz, Max. 0.7A
Display 8.4 inch large high resolution color display,
800x600 pixels
ac40-specs
 

Safety standards

  • IEC60601-1:2005
  • ES60601-1:2005/A2:2010
  • CAN/CSA-C22.2 No. 60601-1:2008
  • IEC60601-1:1988+A1+A2

Classifications

  • Class I. Type B Applied parts

Audiometer standards

  • Tone: IEC 60645-1:2012/ANSI S3.6:2010 Type 1-
  • Speech: IEC 60645-2:1993/ANSI S3.6:2010 Type A or A-E

Calibration

  • ANSI S3.6, ISO 389-1 TDH type;
  • ISO 389-2 Insert;
  • ISO 389-3 Bone;
  • ISO 389-4 NB;
  • ISO 389-5 HF;
  • ISO 389-7 FF angle ISO 8253-2;
  • ISO 389-8 HDA200 LF

Extended range function

  • If not activated, the AC output will be limited to 20 dB below maximum output

Input

  • Tone, Warble Tone, NB, Pediatric Noise, CD1, CD2, Mic1, Mic2, wave files

Masking stimulus

  • Automatic selection of narrow band noise (or white noise) for tone presentation and speech noise for
    speech presentation

Outputs

  • AC Left + Right, BC Left + Right, Insert Phone Left + Right, Insert Masking, FF1+FF2+FF3+FF4, HF Left
    + Right, HLS, Monitor headset. Assist Monitor

Transducers

  • DD45, TDH39, HDA200, HDA280 Audiometric headsets, B71 Bone conductor, EARTone3A/5A insert
    phones

Tone presentation

  • Two patient response push buttons

Patient signal

  • Two patient response push buttons

Intensity

  • AC: -10 to 120 dB HL in 1, 2 or 5dB steps. BC: -10 to 80 dB in 1, 2 or 5 dB steps

Frequency range

  • 125Hz to 8kHz (Optional High Frequency: 8 kHz to 20 kHz - only with HDA200)

Frequency selection

  • 125Hz, 250Hz, 750Hz, 1500Hz and 8kHz may freely be deselected

Patient communication

  • Talk Forward and Talk Back

Monitor

  • Real stereo output through built-in speakers or through external earphone or assistant monitor

Internal storage

  • Internal storage for 500 patients/50,000 test sessions.
  • Multiple user settings (protocols)

AC40 comes with different licence-based test configurations. The test battery includes

  • Speech from hard drive (wave files), ABLB, Stenger, Stenger Speech, Auto threshold, SISI, Langenbeck (tone in noise), Békésy Test, 2 channel tone/speech, 2 channel Master Hearing Aid, QuickSIN™, MLD, Pediatric Noise Stimuli, Multi Frequency, High Frequency, Hearing Loss Simulator

Direct print (USB)

  • Supported printers: Please contact local distributor for a list of approved PC printers.

External devices

  • Standard PC mouse and keyboard (for data entry)

Compatible software

  • Diagnostic Suite - Noah, OtoAccess and XML compatible

Medical CE-mark

  • Yes

Included parts

  • Goose neck microphone
  • DD45 Audiometric headset
  • B71 bone conductor headset
  • Monitor Headset with boom microphone
  • 2 x APS3 Patient response switch
  • Power cable
  • Multilingual CE instructions for use

Optional parts

  • HDA200 High Frequency headset 125Hz to 20 kHz
  • TDH39AA with Amplivox Headset
  • Eartone 5A insert phones
  • Eartone 3A 10 insert phones
  • CIR33 insert earphone kit for masking or monitoring
  • Amplivox audiocups, noise reducing headset
  • Talk back microphone
  • Sound field speakers SP90 (with external power amp)
  • AP70 Power Amplifier 2x70 Watt
  • Cable USB 2m
  • Diagnostic Suite
  • OtoAccess™ database
You are here: Home AC40 Specifications
22 May 2013

PC-based Audiometers

Screening Audiometers

Hybrid Audiometers

Standalone Audiometers

Integration software

Support

Go to the main e-iconSupport section

Events